ALPHATEC SPINE INC ANTERIOR CERVICAL PLATE ASSEMBLY, 4-LEVEL, 068 MM (TI-6AI-4V ELI, NITINOL); SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS Back to Search Results
Model Number 71004-068
Device Problems Break (1069); Detachment Of Device Component (1104); Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2016
Event Type  malfunction  
Manufacturer Narrative
Review of the device history records revealed no manufacturing, processing or design related irregularities.The trestle luxe plate was found to be properly manufactured and released in accordance with the device master record.Visual examination of the returned 4-level plate found that the middle blocking slide had become completely detached from the implant.Witness marks along both the anterior and posterior sides are present and reveal the location of the instrument to plate during contouring/bending.The plate contains a witness mark on the posterior side directly behind the location of the detached blocking slide and witness marks at both graft windows on the anterior side.The combination of these witness marks confirms the plate bending instrument was positioned inconsistent with ifu requirements while contouring/bending the plate.Alphatec surgical technique (lit-84315) provides instructions as to proper plate contouring.It also provides caution to reduce/eliminate the risk of this type of event.Caution: do not place the anvil portion of the plate bender over the blocking slide as damage can occur and affect its function.The provided instructions for use (ins-048) state; intraoperative management: trestle luxe anterior cervical plates are contoured to closely match the anatomical configuration of the spine.If the plate cannot be fitted and additional contouring is necessary, it is recommended that such contouring be minimal and be performed with the instrumentation provided.The plate must not be contoured in proximity of bone screw pockets or screw retention mechanism.When contouring the plate, great care should be taken not to scratch, notch or dent the surface as such deformities may compromise the strength of the implant.
 
Event Description
The locking slide broke off two separate 4-level trestle luxe plates while bending.The second plate was implanted but a screw/anchor was not inserted at the location where the locking slide was missing.The event caused no patient issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTERIOR CERVICAL PLATE ASSEMBLY, 4-LEVEL, 068 MM (TI-6AI-4V ELI, NITINOL)
Type of Device
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS
Manufacturer (Section D)
ALPHATEC SPINE INC
5818 el camino real
carlsbad CA 92008
Manufacturer Contact
richard welch
5818 el camino real
carlsbad, CA 92008
7604946611
MDR Report Key5796308
MDR Text Key50520110
Report Number2027467-2016-00046
Device Sequence Number1
Product Code KWQ
UDI-Device Identifier00844856067196
UDI-Public00844856067196
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number71004-068
Device Catalogue Number71004-068
Device Lot Number694230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-