Catalog Number EMAX2PLUS |
Device Problems
Device Inoperable (1663); Device Displays Incorrect Message (2591); Moisture or Humidity Problem (2986)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper cleaning and maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was observed that the motor device had no rotation and was not functioning.It was noted that the motor stopped working because there was water inside the engine.It was also noted that during pre-repair diagnostic testing the device displayed an e6 error code as soon as it was connected.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|