The patient was undergoing a coil embolization procedure for a gastrointestinal (gi) bleed using ruby coils.During the procedure, while attempting to reposition the last ruby coil into the target vessel, the physician was advancing and retracting the coil, then realized that the coil had unintentionally detached inside the patient.Therefore, the physician required a snare device to remove the detached ruby coil from the celiac artery and the aorta.Subsequently, approximately ninety percent of the coil was removed and approximately three centimeters of residual coil remained in the common hepatic artery.It was reported that the remaining residual coil was likely of no clinical significance to the patient.The procedure ended at this point with no additional coils placed.There was no report of an adverse effect to the patient.Following the procedure, the patient did well.
|
Results: the pet lock was intact on the proximal end of the ruby coil pusher assembly; the stretch resistant wire (sr wire) was fractured and the embolization coil was detached from the pusher assembly.The proximal constraint sphere was inside the distal detachment tip (ddt).The introducer sheath was loaded incorrectly on the pusher assembly.Conclusions: evaluation of the returned device revealed that the ruby coil¿s sr wire was fractured.This type of damage typically occurs due to improper handling during use.If the ruby coil is forcefully retracted during repositioning against resistance, damage such as this may occur.The introducer sheath loaded incorrectly on the pusher assembly was likely incidental and may have occurred during packaging of the device for return.All penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|