Model Number 3100B |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/21/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Should any additional information be received by the customer then an additional report will be submitted.(b)(4).At this time, carefusion has not received the suspect device/component from the customer for evaluation.
|
|
Event Description
|
The customer reported that this 3100b high frequency oscillating ventilator (hfov) was being used on a patient when the driver unexpectedly stopped.The device alarmed and the hfov was removed from the patient and alternate ventilation was provided by placing a known good unit on the patient.After the suspect unit was removed from the patient, the customer reported that the unit failed the patient circuit calibration.The customer stated that there was no patient injury or harm associated with this event.
|
|
Manufacturer Narrative
|
The 3100b ventilator was due for the 6 year and 7 year preventative maintenance.A carefusion field service representative (fsr) arrived onsite to evaluate the device, however, the customer opted to not have the preventative maintenance procedures performed on the unit.
|
|