The device was not returned to olympus for evaluation.The facility uses a third party repair entity to repair and service the scope.The root cause of the reported event cannot be determined.The instruction manual warns users: ¿confirm that the slit and hole on the biopsy valve have no splits, cracks, deformations, discoloration, or other damage.The biopsy valve is a consumable that should be inspected as follows before each use.Replace it with a new one if any irregularity is observed during the inspection.¿.
|