HOLOGIC, INC ATEC INTRODUCER LOCALIZATION SET SURESIGHT; SURGICAL NEEDLE GUIDE Back to Search Results
Model Number 0914-20-OB
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/23/2016
Event Type  malfunction  
Manufacturer Narrative
The device has not yet been returned therefore, a failure analysis of the complaint device cannot be completed.If the device is returned and evaluation completed, a supplemental medwatch will be filed.Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Currently unable to establish a relationship or impact to the reported observation.(b)(4).
 
Event Description
It was reported during a breast biopsy procedure (exact date unknown) the obturator tip fell off outside the patient prior to insertion.No report of patient injury.
 
Manufacturer Narrative
The returned device was received and visually inspected.The investigator verified that the distal tip of the obturator is broken.Cause is presently unknown for failure.The reported observation was confirmed.This observation is being trended and monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATEC INTRODUCER LOCALIZATION SET SURESIGHT
Type of Device
SURGICAL NEEDLE GUIDE
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
sidra piracha
250 campus drive
marlborough, MA 01752
5082638884
MDR Report Key5813092
MDR Text Key51044520
Report Number1222780-2016-00182
Device Sequence Number1
Product Code GDF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/19/2016
Device Model Number0914-20-OB
Device Catalogue NumberILS 0914-20-OB
Device Lot Number602503
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/15/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-