It was reported that the tracheostomy cuff would not inflate and it leaked.The old tracheostomy tube was replaced with the defective tube as part of a weekly routine change.When it was inserted, the ventilator started to alarm.The patient's mother removed the tube and tried to inflate it, but it would not inflate.Sterile water would not go to the balloon from the cuff.It was unknown if cuff patency was checked prior to use.The tube was switched out with another tube in a non-urgent matter.There was no injury associated with the issue.
|
A 3.5mm tts pediatric flextend trach was returned for cuff leak.A visual inspection was performed and a small cut was observed on the balloon area.The cut measured 1/2mm in length.Functional testing was performed using a standard syringe, 5cc's of air was pushed through the inflation line; the cuff did not inflate.System leak test was performed on the device by placing in a pan of water, pushing air through the inflation line, water was observed coming from the balloon.The instruction for use (ifu), states the following: do not use a tube that is cut or damaged.Use of a damaged tube can result in airway compromise.This device must be thoroughly inspected for signs of damage or wear prior to each use.Guard against product damage by avoiding contact with sharp edges.The ifu also recommends the use of twill.The type of trach holder identified in the complaint description, contains a ridged velcro strap that has the potential to cut the eyelet promoting a tear.Root cause is the failure as a result of the customer/ user interfacing with the product in a manner inconsistent with the ifu.(b)(4).
|