Model Number 840
Device Problems Thermal Decomposition of Device (1071); Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Date 08/28/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the ventilator had a burning smell while calibrating the oxygen sensor.The graphical user interface (gui) printed circuit board (pcb) was replaced.The gui pcb was returned for further investigation.A visual inspection was performed on the returned gui pcb.Server thermal damaged was observed one of the capacitor and the immediate pcb surrounding area.The possible causes of the type of component failure are electrical overstress.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hampshire st.
mansfield, MA 02048
5084524811
MDR Report Key5820977
MDR Text Key51056499
Report Number8020893-2016-01646
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number4-840220DIUA-EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-