(b)(4).The opt944 interface is used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint cannulae were not returned to fisher & paykel healthcare for evaluation.Our investigation is based on a photograph of one of the complaint devices provided by the customer.Results: visual inspection of the photograph revealed that the tubing appeared to have been pulled out of the swivel connector.Conclusion: the observed damage is most likely the result of pulling on the cannula tubing with undue force.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and stretching or deformation.Any product that fails the visual inspection is disposed of.The setup instructions in the user instructions which accompany the opt944 nasal cannula include the following steps: ensure head strap clip is attached, to prevent cannula from being pulled out of the nares; cannula can become unattached if not used with the head strap clip; attach tubing clip to clothing/bedding to prevent cannula from pulling off face.The user instructions also contain the following warnings/cautions: do not crush or stretch tube, to prevent loss of therapy; failure to use the set-up described above can compromise performance and affect patient safety.Our optiflow production team has been notified of this failure.They have carried out an internal investigation of the manufacturing process and production staff have received additional training to ensure they are following the correct assembly procedure.
|