DEPUY SYNTHES POWER TOOLS MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT
|
Back to Search Results |
|
Catalog Number MA-D20 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported that during a laminectomy surgical procedure, it was observed that the attachment device heated up.According to the report, the surgeon was holding the attachment at the elbow of the device and about 10 seconds into the instrument assembly check, the surgeon asked for a backup due to the attachment heating up to an uncomfortable level.It was further determined that the device had not yet been used on the patient.There were no reports of any delays in the surgical procedure.It was reported that a spare device was available for use.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition that the device was heating up was not confirmed.Therefore, an assignable root cause could not be determined.However, during evaluation, it was determined that the bearings and gears were wornout and broken causing the device to not rotate.It was determined that these issues were due to worn out bearings from normal use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|