Model Number SDM-00001
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/27/2016
Event Type  malfunction  
Manufacturer Narrative
The c-arm hardware was loose with a lot of play in the meshing of the rotational gears.All hardware was tightened, which corrected the problem.
 
Event Description
The c-arm was jerky as the arc went across patient as rotated around, with patient in compression.There was no affect image quality, nor was there any injury to the patient.This has been an intermittent problem for a couple of weeks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
debra rosenberg
36 & 37 apple ridge road
danbury, CT 06810
2032074512
MDR Report Key5828128
MDR Text Key50616445
Report Number1220894-2016-00022
Device Sequence Number1
Product Code OTE
UDI-Device Identifier15420045500013
UDI-Public15420045500013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PO10025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial
Report Date 07/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberSDM-00001
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-