OSCOR INC. MEDTRONIC REUSABLE EXTENSION CABLE; ELECTRODE CABLE Back to Search Results
Model Number 53912
Device Problem Pacing Problem (1439)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2016
Event Type  malfunction  
Manufacturer Narrative
The device was in use for treatment.The device was not returned for evaluation; as a result, the allegations against this cable (no output) cannot be confirmed.A review of the device history records could not be performed as the lot number of the device is unknown.As the device was not returned for analysis, the root cause of the failure could not be determined.Potential cause of this failure: insufficient strength of soldered joint.Since the separation between the cable wire and the soldered pin may cause a disruption of the cable function, a corrective and preventive action has been opened to address this failure.Per the cable inspection procedure, during in process inspection the cable is inspected 100% for continuity.During final inspection, each strand of the cable is checked for the correct colored part, overall length of the cable, cable checked for damage and insulation voids, verification that the silicone strain relief extends at least for a length of 2 cm outside the female connector, checked that the silicone strain relief fits the wire snugly, checks for strain relieves for damage and gaps, a pull test is done on each connector and pin and the cable is inspected again for continuity.The instructions for use (ifu) informs the user: for reusable cables only: the cables can be re-sterilized by oscor eto gas sterilization a maximum of two times.The directions for use inform the user to: first attach the proximal connector(s) to a pacemaker/analyzer or another extension cable while observing polarity indicated by the color.Then attach the distal connector(s) to the temporary lead connector or to another extension cable, again observing polarity by the color.The detachment of the connection should be done in the reverse order.Precautions include: do not connect any cable model to a/c power source; connection of exposed pins to a/c power source may pose a risk of serious injury or death.Extension cables are not intended for use with apnea monitors.
 
Event Description
The customer reported the external pacer 5391 was not firing.Biomed tested on qed and was able to measure an output, however when using the adapter cable 53912 there was no output.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC REUSABLE EXTENSION CABLE
Type of Device
ELECTRODE CABLE
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer Contact
dorit segal
3816 desoto blvd.
palm harbor, FL 34683-1618
7279372511
MDR Report Key5837911
MDR Text Key50915133
Report Number1035166-2016-00139
Device Sequence Number1
Product Code IKD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070926
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial
Report Date 07/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number53912
Device Catalogue Number53912
Is the Reporter a Health Professional? No
Date Manufacturer Received07/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-