The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The tube sheath was kinked.As the result of checking the manufacturing record of the same lot, there was nothing abnormal detected.Based on the evaluation of the subject device and the similar cases in the past, the retraction failure of the needle tube might be caused by increasing of frictional resistance between the tube sheath and the needle tube.Increasing of frictional resistance might be caused by kinking of the tube sheath.The kinking might be caused by excessive load given to the subject device due to inserting and withdrawing it to an endoscope vigorously.The instruction manual of the subject device warns; *when inserting the instrument into the endoscope, retract the needle into the tube, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.*insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.*stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|
During a colorectal polypectomy, the doctor pulled the slider of the subject device and withdrew the subject device from the endoscope.Since the needle tube of the subject device was not retracted into the tube sheath, the needle tube of the subject device stuck to the nurse who operated the subject device.It was reported that the nurse was not given a medical treatment.The doctor completed the procedure with the subject device.
|