Siemens filed the initial mdr 1219913-2016-00130 on august 02, 2016.On 08/16/2016 additional information: a siemens field service engineer (fse) was sent to the customer site.The instrument performance was tested.Only the acid pump and valve were replaced as a countermeasure.Alignment for probes 1,2, and 3 were performed.Dispense test for acid was performed.No issues were found.The patient sample is not available for further testing and investigation.The cause for the discordant hav igm and hav total results is unknown.The instrument is performing within specifications.No further evaluation of the device is required.
|