Results: the pet lock was intact.The ruby coil pusher assembly was kinked approximately 5.0, 30.0, 54.0, and 107.0 cm from the proximal end.The coil was attached to the distal detachment tip (ddt), and compressed away from the proximal constraint sphere.The coil was damaged throughout the length.Conclusions: evaluation of the returned device revealed that the coil was compressed and the overall coil structure was compromised.This type of damage typically occurs due to improper handling during use.If the ruby coil is manipulated with force at an angle, it is likely that this type of damage could occur.Despite the structural damage to the coil, the coil shaped properly when introduced into body temperature water.The root cause of this complaint could not be determined.These devices are 100% functionally tested and visually inspected during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing an embolization procedure using ruby coils.During the procedure, while the physician was deploying a ruby coil in the patient, the coil would not shape properly in the target vessel.Therefore, the ruby coil was withdrawn through the px slim delivery catheter (px slim) and the procedure was completed using a new ruby coil and the same px slim without any further issues.There was no report of any adverse event on the patient.
|