Model Number MODEL 250 |
Device Problems
Failure to Interrogate (1332); Defective Device (2588)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported that the tablet could not complete interrogation of generators.Troubleshooting was performed and the connection of the usb cable to the tablet was suspected to be at fault.The nurse reported that when the usb cable was replaced, the issue was resolved.No additional information was provided to date.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
|
|
Manufacturer Narrative
|
Describe event or problem; corrected data: the previously submitted mdr inadvertently reported a usb cable issue already reported in other medwatches.
|
|
Event Description
|
Further clarification was received that the usb cable issue reported in the present medwatch is a duplicate of one of the tree reported in the following medwatch: mfr.Report # 1644487-2016-01770, mfr.Report # 1644487-2016-01771, mfr.Report # 1644487-2016-01772.It was reported that all tree programming computers were involved.
|
|