Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that the tablet could not complete interrogation of generators.Troubleshooting was performed and the connection of the usb cable to the tablet was suspected to be at fault.The nurse reported that when the usb cable was replaced, the issue was resolved.No additional information was provided to date.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
 
Manufacturer Narrative
Describe event or problem; corrected data: the previously submitted mdr inadvertently reported a usb cable issue already reported in other medwatches.
 
Event Description
Further clarification was received that the usb cable issue reported in the present medwatch is a duplicate of one of the tree reported in the following medwatch: mfr.Report # 1644487-2016-01770, mfr.Report # 1644487-2016-01771, mfr.Report # 1644487-2016-01772.It was reported that all tree programming computers were involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5853267
MDR Text Key52060176
Report Number1644487-2016-01773
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/09/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-