• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Defective Device (2588); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported than a medical professional could not interrogate generators when using her programming system.Troubleshooting was performed and the connection of the usb cable to the tablet was suspected to be at fault.Review of manufacturing records confirmed that the tablet passed all functional tests prior to distribution.The suspected usb cable was returned to the manufacturer.Analysis is underway but it has not been completed to date.
 
Event Description
An analysis was performed on the returned usb to db9 cable and the reported allegation was verified.The cause for the reported allegation is associated with a disconnected wire connection in the returned serial cable.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5853315
MDR Text Key52061397
Report Number1644487-2016-01774
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Assistant
Type of Report Initial,Followup
Report Date 07/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-