One catheter with attached monoject 1.5 cc limited volume syringe and two three-way stopcocks was returned for evaluation.A non-edwards introducer with non-edwards contamination shield was located on the catheter body between 7 cm and 98 cm proximal from the catheter tip.No packaging was returned.As received, blood was observed from inside the attached non-edwards introducer tube, under the balloon latex, and inside the gate valve.The attached non-edwards introducer and non-edwards contamination shield were removed for further evaluation.Two punctures, were found on the catheter body at 5.5 cm proximal from the catheter tip.The punctures measured, <0.5mm long and <0.5 mm long, respectively and were located on the opposite side of each other.The punctures entered into the distal lumen and balloon inflation lumen.The proximal injectate lumen was patent without any leakage or occlusion.The balloon did not inflate due to clotted blood occlusion.No visible damage to the balloon or returned syringe was observed.Balloon inflation test was performed using lab syringe with 1.5 cc air.Visual examination was performed under microscope at magnification 20x and with the unaided eyes.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Customer report of ¿the catheter could not wedge in pulmonary artery¿ could not be confirmed during evaluation, however, blood leakage issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.
|