Catalog Number 82090-01
Device Problem Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the procedure was to treat a lesion in the carotid artery with 95% stenosis.An 8x40mm xact stent system met resistance when advancing over an emboshield bare wire.The stent system was withdrawn with resistance.A new same size xact was able to traverse the bare wire and was successfully used to treat the lesion.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: visual and functional inspection was performed on the returned device.The difficulties were not confirmed.The investigation was unable to determine a conclusive cause for the reported difficulties.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Type of Device
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5857833
MDR Text Key51923708
Report Number2024168-2016-05267
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue Number82090-01
Device Lot Number5091661
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/05/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
Patient Weight83
-
-