• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICROBLENDER
Device Problems False Alarm (1013); No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 07/13/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).At this time, carefusion has not received the suspect device/component from the customer for evaluation.In the event the device is received for evaluation or additional information is received a follow-up report will be submitted.
 
Event Description
The customer stated that this blender was not alarming at all and after replacing the alarm reed, o-rings, and poppet it is alarming constantly.There was no patient involvement at the time of the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
7149227830
MDR Report Key5861054
MDR Text Key51818372
Report Number2021710-2016-04203
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMICROBLENDER
Device Catalogue Number10065
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-