Model Number TABLETOP
Device Problems Device Emits Odor (1425); Loss of Power (1475); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/25/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A hospital service technician reported that a laser system presented with a system message, burnt smell and a laser shut down during a procedure.The procedure was completed.There was no patient harm.
 
Manufacturer Narrative
The system was examined and the reported events were replicated.The printed circuit board (pcb) was replaced to address the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on february 22, 2010.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the pcb.However, how or when the component became nonconforming cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key5866545
MDR Text Key52507328
Report Number2028159-2016-03496
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Hospital Service Technician
Type of Report Initial,Followup
Report Date 09/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2020
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number2.02.60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-