• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRICURE, INC. COBRA FUSION 150 ABLATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRICURE, INC. COBRA FUSION 150 ABLATION SYSTEM Back to Search Results
Model Number 001-700-001S
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received and sent to atricure engineering for analysis.The complaint was confirmed device was not functioning properly.Electrodes have been dislodged and system generates error code when ablating on the bench.
 
Event Description
It was reported during ablation, electrode 2 and 3 didn't functioned normally, in mono or in bi polar.Electrode 1 was okay and vacuum was okay.The surgeon moved the probe many times to perform correct ablation only with electrodes.One, 4, 5 and 6.The case was completed and the patient outcome was not affected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBRA FUSION 150 ABLATION SYSTEM
Type of Device
COBRA FUSION 150 ABLATION SYSTEM
Manufacturer (Section D)
ATRICURE, INC.
6217 centre park drive
west chester OH 45069 3886
Manufacturer Contact
ranjana iyer
6217 centre park drive
west chester, OH 45069-3886
5137555320
MDR Report Key5867010
MDR Text Key51887899
Report Number3003502395-2016-00085
Device Sequence Number1
Product Code OCL
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K113475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 01/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date07/01/2017
Device Model Number001-700-001S
Device Catalogue Number001-700-001S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/15/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-