The patient was undergoing a microneurosurgery using the apollo system.During preparation for the procedure, it was noticed that the apollo canister cup holder (holder) bracket was broken on one side.The hospital staff secured the holder with tape and it worked fine for the procedure.During the procedure, upon initiating aspiration and irrigation, it was noticed that the peristaltic pump clamp on the apollo generator (generator) was not working properly, and failed to properly secure the apollo irrigation tubing (tubing).The tubing remained secure from advancing by the pump and worked fine.There was minimal time delay as the pump and tubing continued to work throughout the procedure despite the tubing connector advancing too close to the pump.The procedure was completed using the same apollo system.There was no report of an adverse effect to the patient.
|
Results: the canister holder tab was broken.Conclusions: evaluation revealed that the canister holder was broken.This type of damage typically occurs due to improper handling.If the canister was overtightened while attaching the canister holder to the apollo system, damage such as this may occur.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.
|