• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SCD 700 COMPRESSION SYSTEM-US; SCD CONTROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SCD 700 COMPRESSION SYSTEM-US; SCD CONTROLLER Back to Search Results
Model Number 29525
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 08/12/2016.An investigation is currently underway.Upon completion, a full detailed investigation will be provided.
 
Event Description
On (b)(6) 2016 the customer states the unit has a broken screen.Upon triage on (b)(6) 2016 the service tech found the unit had a damaged power cord with exposed copper wire.
 
Manufacturer Narrative
A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; a broken screen with a service finding of a damaged power cord with exposed copper wire.Therefore, this report will be based on information provided by the technical center.The scd 700 was evaluated and the customer reported issue was confirmed; the power cord was confirmed to have exposed copper wire.The cause of the reported condition was determined to be due to handling.The power cord was replaced and scrapped to correct the reported issue.The unit passed all initial testing after failure analysis repair instructions were completed.Scd 700 was manufactured in 2013.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD 700 COMPRESSION SYSTEM-US
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
Manufacturer (Section G)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH  
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02448
5084524151
MDR Report Key5871370
MDR Text Key51949313
Report Number3006451981-2016-00327
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number29525
Device Catalogue Number29525
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/13/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-