Catalog Number 544965 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/27/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device investigation has not been submitted at this time.The manufacturer will continue to monitor and trend related events.
|
|
Event Description
|
The clips fell from applier easily and some clips were dropped in the patient during the procedure.All the clips were retrieved.The patient's condition was reported as fine.
|
|
Manufacturer Narrative
|
(b)(4).The dhr for the instrument in question, was reviewed and found completely without any irregularities.This instrument was manufactured (b)(4) in may of 2015.The returned instrument was evaluated and found that the jaws are aligned with each other and the tip set as received is correct per print specs.Further evaluation showed that this instrument picks up, retains, closes and releases clips both with and without the use of silastic test tubing as required of its function.We are unable to validate the alleged complaint since we are unable to duplicate the alleged issue.Parts were 100% visually inspected and tested at the (b)(4) facility before instruments were sent to customer.No irregularities were found and or reported at the time of inspection and assembly of the product as this is a standardized process for all instruments manufactured at this facility.No corrective action required at this time.
|
|
Event Description
|
The clips fell from applier easily and some clips were dropped in the patient during the procedure.All the clips were retrieved.The patient's condition was reported as fine.
|
|