Model Number ETH2B5LT |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not yet been returned to the manufacturer at the time of this report.A supplemental form will be sent once the evaluation is completed if the device is returned.The device history report was reviewed and no discrepancies were found.
|
|
Event Description
|
It was reported that during a procedure, debris was found in the device.It is reported that the patient was being monitored for adverse reaction, but at the time of this report, no patient harm or consequences were noted.
|
|
Manufacturer Narrative
|
The account supplied photos showing the debris inside the obturator, along with the debris missing from the camera view in the field.When the device was returned, the rear port was uncovered and no debris was found to be present.The debris in the photos supplied by the account is reddish/brown which is inconsistent with organic debris that has been exposed to eo sterilization.The photos supplied by the account confirm that debris was present during use in the field, though no analysis of the debris could be completed due to the debris being missing when returned.
|
|