Model Number 3116
Device Problems Failure to Deliver Energy (1211); Device Displays Incorrect Message (2591); Battery Problem (2885); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
The manufacturer representative reported that twice the patient's implantable neurostimulator (ins) lost the serial number and turned off.The device was interrogated and kept resetting itself.It was reviewed that this was because the ins had a power on reset (por) and reset.The most likely cause was a low ins battery.The ins needed to be replaced for the low battery.The patient was scheduled to have the battery changed.The health care provider (hcp) saw the patient last week for the issue.The por and low battery on the ins had not been resolved yet.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5879719
MDR Text Key52232007
Report Number3004209178-2016-16930
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2013
Device Model Number3116
Device Catalogue Number3116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/25/2016
Date Device Manufactured10/26/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-