Catalog Number J-CHSG-703001 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/20/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The event is currently under investigation.
|
|
Event Description
|
During the procedure, the cook silicone balloon hysterosalpingography injection catheter was inflated successfully; however when it was time for removal the balloon would not deflate.The user was able to remove the balloon by cutting the port at the bottom.The user elected to open a 2nd cook silicone balloon hysterosalpingography injection catheter to test it before using it before continuing on with the procedure; however the very same difficulty occurred.The procedure was completed with the 1st complaint device as this was towards the end of the procedure.A section of the device did not remain inside the patient's body.According to the initial reporter, the patient did not require any additional procedures nor experience any adverse effects due to this occurrence.
|
|
Manufacturer Narrative
|
Investigation - evaluation: a cook silicone balloon hysterosalpingography injection catheter was not returned/received for an evaluation.The customer report that the balloon would not deflate could only be confirmed based on the customer¿s testimony.Based on the information available a definitive root cause could not be established.A review of production and quality documentation did not identify any specific issues with current manufacturing or quality controls that may have contributed to this incident.A review of the device history record could not be performed due to the lot number was not provided.A review of complaint history for the associated lot could not be performed due to the lot number was not provided.Measures have been initiated to address this failure mode.Per the quality engineering risk assessment; no further action is required.Cook medical has notified the appropriate personnel and the will continue to monitor this device via the complaints database for similar complaints.
|
|
Search Alerts/Recalls
|