(b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The reported device was returned to livanova (b)(4) for further investigation.During visual inspection, a damaged front panel was identified.However, the issue was not reproduced during functional testing.The calibration was renewed and a technical safety inspection was successfully carried out.The unit was returned to the customer.As the issue was not reproduced, a root cause could not be determined.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|