(b)(4).Method: four parts of the rt380 adult dual-heated evaqua2 breathing circuits were returned to fph in (b)(6) for evaluation: the elbow connector and part of the evaqua tube (device 1), the patient end connector and part of the evaqua tube (device 2), the elbow connector and part of the evaqua tube (device 3) and a part of the evaqua tube (device 4).The subject breathing circuits were visually inspected.Results: visual inspection of the complaint devices revealed that the expiratory elbow were found cracked (device 1 and device 3), the patient end connector were found cracked and the evaqua tube was found to be brittle and degraded (device 2) and the evaqua tube was found to be brittle and degraded (device 4).A lot check was not performed as lot information was not provided.Conclusion: the hospital informed fph that the hospital is disinfecting breathing circuit systems before use.It is possible that the affected breathing circuits also came into contact with a chemical solution, resulting in environmental stress cracking of the connectors and leading to a degradation of the expiratory limbs.All rt380 adult breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject rt380 adult breathing circuits would have met the required specification at the time of production.The user instructions that accompany the rt380 adult dual heated evaqua2 breathing circuit state: do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
|