• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON TUBING, OXYGEN SUPPLY 7'; PRESSURE TUBING & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON TUBING, OXYGEN SUPPLY 7'; PRESSURE TUBING & ACCESSORIES Back to Search Results
Catalog Number 1115
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device sample was not returned for evaluation at the time of this report.
 
Event Description
The customer alleges that the oxygen supply tubing kinked during use.No patient injury reported.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no kinking was observed on the tubing.Functional testing was also performed and no issues were found.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
The customer alleges that the oxygen supply tubing kinked during use.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON TUBING, OXYGEN SUPPLY 7'
Type of Device
PRESSURE TUBING & ACCESSORIES
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5885800
MDR Text Key52414035
Report Number3004365956-2016-00304
Device Sequence Number1
Product Code BYX
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number1115
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-