• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY VDW GMBH ENDO IT PROFESSIONAL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY VDW GMBH ENDO IT PROFESSIONAL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Device Problems Break (1069); Device Damaged by Another Device (2915)
Patient Problems No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Event Type  malfunction  
Manufacturer Narrative
Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The file separation events will be reported via asr, as appropriate.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that an endo it control motor was involved in two file breakages.One file could be retrieved, while the other was incorporated into the filling.
 
Manufacturer Narrative
Evaluation found the bearings were damaged.Replaced with new motor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO IT PROFESSIONAL
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
Manufacturer (Section G)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM   81737
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5886985
MDR Text Key53394096
Report Number9611053-2016-00039
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 07/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-