STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMP #8R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5515F802
Device Problems Device Damaged Prior to Use (2284); Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 07/20/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Packaging was damaged.Plastic package containing the implant was cracked.No noticeable damage to the outer box or plastic wrap on outside of box.
 
Manufacturer Narrative
An event regarding packaging damage involving a triathlon femoral component was reported.The event was confirmed.Device evaluation and results: based on the visual inspection of the returned packaging it appears that this component packaging was subjected to excessive/inappropriate handling during transportation/storage whereby the carton had been compressed.Medical records received and evaluation: not performed as the event is related to a packaging issue and no adverse consequences to the patient were reported.Device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: there have been no other events for the lot referenced.The investigation concluded that the packaging damage was most likely caused by excessive/incorrect handling prior to the opening packaging.No further investigation for this event is required at this time.
 
Event Description
Packaging was damaged.Plastic package containing the implant was cracked.No noticeable damage to the outer box or plastic wrap on outside of box.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMP #8R-CEM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5887042
MDR Text Key53119655
Report Number0002249697-2016-02649
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Catalogue Number5515F802
Device Lot NumberECTSF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-