Catalog Number 03.812.003
Device Problem Dull, Blunt (2407)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/05/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Patient weight is unknown.Additional product codes for this report include lxh.(b)(4) lot unknown.Device is an instrument and is not implanted or explanted.The complainant part is not expected to be returned for manufacturer review/investigation, but has yet to be received.Investigation could not be completed and no conclusion could be drawn as no device was returned.Without a lot number, the device history record review could not be requested.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a patient underwent a transforaminal lumbar interbody fusion (tlif) procedure from l4-s1 on (b)(6) 2016 using a transforaminal posterior atraumatic lumbar (t-pal) system.As the implant was being loaded, the push release button on the t-pal spacer applicator knob became jammed and the proximal end of the t-pal spacer applicator inner shaft became stripped.Secondary instrumentation was used to complete the procedure with a twenty (20) to thirty (30) minute delay.The patient's post-operative status was reported as stable.Concomitant devices reported: t-pal spacer applicator handle (part: 03.812.001 / lot: unknown / quantity: 1).This report is 2 of 2 for (b)(4).
 
Manufacturer Narrative
Subject device has been received and is currently in the evaluation process.Investigation is ongoing; no conclusion could be drawn as product is entering the complaint system.Device history records review was conducted.The report indicates that the: dhr review for: part # 03.812.003, lot # 9428484, manufacturing site: haegendorf, manufacturing date: 19.May 2015.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device used for treatment, not diagnosis.If in formation is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis a product investigation was completed: the returned inner shaft was examined and the complaint condition was able to be confirmed as the proximal coupling head was found to be broken and retained by the returned application knob (03.812.004).The remainder of the device was found to be intact with minimal witness marks concentrated on the distal prongs/spacer consistent with wear and tear; these marks would not inhibit device functionality.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.A visual inspection, functional test, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.No product design issues or discrepancies were observed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL SPACER APPLICATOR INNER SHAFT
Type of Device
INTERVERTEBAL FUSION DEVICE W/BONE GRAFT, LUMBAR
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf PA CH461 4
SZ  CH4614
Manufacturer (Section G)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH461 4
SZ   CH4614
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5888354
MDR Text Key52496115
Report Number2520274-2016-14221
Device Sequence Number1
Product Code MAX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151276
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.812.003
Device Lot Number9428484
Other Device ID Number(01)10705034727790(10)9428484
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
-
-