Catalog Number 530.610
Device Problems Vibration (1674); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition of the device not oscillating was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device emitted an unusual noise.It was further determined that the electric motor was damaged.It was observed that the device had a rough rotation (vibration) and corrosion.It was further determined that the o-rings were damaged and the bearings were worn.The assignable root cause was determined to be due to component wear from normal use over time.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during pre-surgery, it was discovered that the battery oscillator device would not oscillate.During in-house engineering evaluation, it was observed that the device had a rough rotation (vibration).There were no delays to the surgical procedure as a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
terry callahan
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5888734
MDR Text Key53465367
Report Number8030965-2016-14597
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.610
Device Lot Number4143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2016
Date Manufacturer Received08/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-