• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYER MEDICAL CARE, INC. STELLANT DUAL CT INJECTOR; CT INJECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYER MEDICAL CARE, INC. STELLANT DUAL CT INJECTOR; CT INJECTION SYSTEM Back to Search Results
Model Number 3032458
Device Problems Nonstandard Device (1420); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/04/2016
Event Type  malfunction  
Manufacturer Narrative
Bayer service performed a site visit and replaced the aioc (all-in-one computer) display and the system was restored to normal operation.Bayer radiology product analysis received and examined the returned aioc display.Visual examination found the hard drive power cable insulation was thermally degraded.Our investigation determined that a high current condition within the display module had occurred which caused material deformation and degradation of the surrounding plastic connectors and wiring insulation.On august 3, 2016, bayer medical care distributed a field safety notice recalling affected certegra® workstation all-in-one computers (aioc) with part number 3030896 that are used in conjunction with the medrad® stellant® ct injection system.
 
Event Description
The following information was received from a bayer representative: a customer reported smoke was seen coming from the medrad stellant ct injection system all-in-one computer (aioc) display.No injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT DUAL CT INJECTOR
Type of Device
CT INJECTION SYSTEM
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer (Section G)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
Manufacturer Contact
lois miller
1 bayer drive
indianola, PA 15051
7249407497
MDR Report Key5893523
MDR Text Key53595510
Report Number2520313-2016-00057
Device Sequence Number1
Product Code DXT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082905
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Radiologic Technologist
Remedial Action Recall
Type of Report Initial
Report Date 08/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model Number3032458
Device Catalogue NumberSCT W/CERTEGRA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-