• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 08/01/2016
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.On 08/17/2016 a medtronic representative, following-up at the site, reported the site biomed representative worked on the articulating arm by adjusting the knob.The site is no longer interested in replacing the articulating arm.No parts were replaced.No parts have been received by manufacturer for analysis.No further issues have been reported.
 
Event Description
A medtronic representative reported a site navigation system articulating arm that would not tighten properly.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key5894535
MDR Text Key52739862
Report Number1723170-2016-03092
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 08/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? No
Date Manufacturer Received08/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-