Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.As such, the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
It was reported that a patient underwent an ablation procedure for atrial fibrillation with a pentaray nav eco catheter where the tip of the catheter clotted off and could not be flushed.For the first part of the procedure, the catheter was functioning properly.About halfway through the case, the irrigation pump started to alarm, indicating that the irrigation was no longer functioning.The catheter was removed from the patient, but no clot was visible.A manual flush was not successful, so the catheter was replaced and the case continued without any patient consequence.Ablation was being performed in power control mode with power set to 25-35w, temperatures between 34-43 degrees celsius and impedance between 100-140 ohms.The patient was being administered an anticoagulant during the procedure (heparinized saline, 2u/cc).No errors were noted on the carto system.Blood clots exhibit poor adherence to catheters and transseptal sheaths, making them a potential source of embolism for the patient.As a result, this event is mdr reportable.
|