• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-UNI-CELECT-PT
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/19/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog# igtcfs-65-1-uni-celect-pt (b)(4).Summary of investigational findings: no product or images have been received for investigation, such that the evaluation is based on the description solely.Based on the information provided, the exact reason for the deployment difficulties cannot be determined, however it is known from the literature that excessive back tension could result in deployment difficulties/failure when pressing the release button.From the ifu: - while keeping slight back tension on the introducer push the release button completely to ensure proper release of the filter.Repositioning of the filter is no longer possible.The filter is now released.- excessive tension during deployment may prevent the filter from releasing when release mechanism is activated.There is found no evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to complainant: the filter would not deploy.After several attempts from the jugular approach, the filter would not unhook from the system.The physician re sheathed the filter a few times and then was finally able to detach it from the deployment system.Everything ended up being ok.Patient outcome: the patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key5899579
MDR Text Key52847984
Report Number3002808486-2016-00947
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002345055
UDI-Public(01)10827002345055(17)170903(10)E3251812
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-1-UNI-CELECT-PT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/23/2014
Date Device Manufactured09/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-