• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONFORMIS, INC. IDUO G2; BICOMPARTMENTAL KNEE REPLACEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONFORMIS, INC. IDUO G2; BICOMPARTMENTAL KNEE REPLACEMENT SYSTEM Back to Search Results
Catalog Number M5723INT0600210
Device Problems Break (1069); Detachment Of Device Component (1104); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2016
Event Type  malfunction  
Manufacturer Narrative
The disposable tibial impactor tip became lodged in the disposable impactor handle and broke off in the impactor handle during attempts to remove the tip.The handle could not be used to impact the femoral implant so the femoral implant was impacted with a different instrument.Review of the device history record indicates the device was manufactured to specification.
 
Event Description
The disposable tibial impactor tip became lodged in the disposable impactor handle and broke off in the impactor handle during attempts to remove the tip.The handle could not be used to impact the femoral implant so the femoral implant was impacted with a different instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDUO G2
Type of Device
BICOMPARTMENTAL KNEE REPLACEMENT SYSTEM
Manufacturer (Section D)
CONFORMIS, INC.
28 crosby dr.
bedford MA 01730
Manufacturer (Section G)
CONFORMIS, INC.
600 research dr.
wilmington MA 01887
Manufacturer Contact
karina snow
28 crosby dr.
bedford, MA 01730
7813459195
MDR Report Key5900040
MDR Text Key53707546
Report Number3004153240-2016-00173
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K093513
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/31/2017
Device Catalogue NumberM5723INT0600210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
-
-