• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY, INC. G-FORCE(TM) HEX DRIVER; SMALL JOINT INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY, INC. G-FORCE(TM) HEX DRIVER; SMALL JOINT INSTRUMENT Back to Search Results
Catalog Number 86PS-3016
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2016
Event Type  malfunction  
Manufacturer Narrative
The investigation is not complete.Trends will be evaluated.This report will be updated when the investigation is complete.This is the same event as 1043534-2016-00088.
 
Event Description
Allegedly, when bringing in the implants, the screw driver would not release from the implant.The implant had to be pulled out with the screw driver.No delay was caused.
 
Manufacturer Narrative
This incident is considered closed.If at any time new or updated information becomes available, the incident will be re-opened and investigated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G-FORCE(TM) HEX DRIVER
Type of Device
SMALL JOINT INSTRUMENT
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
1023 cherry road
memphis, TN 38117
901451-631
MDR Report Key5900858
MDR Text Key53842505
Report Number1043534-2016-00087
Device Sequence Number1
Product Code HXX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue Number86PS-3016
Device Lot Number692771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date06/17/2016
Date Manufacturer Received06/17/2016
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage N
Patient Sequence Number1
-
-