• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/28/2016
Event Type  malfunction  
Manufacturer Narrative
Patient id was not available.Evaluation of suspect probe confirmed reported event: the tip was bent.This issue has been documented in a medtronic hardware defect database for trending and monitoring.
 
Event Description
A medtronic representative reported that the pedicle probe became bent during a spine surgery.They were using it, and when they removed it, they noticed it was slightly bent.They used a different instrument for the remainder of the case.There was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
catherine eaton
826 coal creek circle
louisville, CO 80027
7208902092
MDR Report Key5902800
MDR Text Key52953066
Report Number1723170-2016-03124
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NAVIGATED VERTEX® MAX 2.25MM PEDICLE PROBE
Patient Age13 YR
Patient Weight36
-
-