• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-2-UNI-CELECT
Device Problems Difficult to Advance (2920); Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2013
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog # igtcfs-65-2-uni-celect.Similar to device under 510(k) k090140.Summary of investigational findings: investigation is based on event description and returned product.Femoral introducer with loaded filter returned.Red locking mechanism still locked.One secondary leg is severely deformed, two other secondary legs are crossed and slightly bending outwards - probably because the filter was retrieved through the check flo.Since the sheath is not returned, the exact reason for the filter/introducer to stick cannot be determined, but the sheath may not be the reason, since a second celect filter could be advanced.Or advancement may have been abandoned, if sheath was kinked and filter "didn't advance normally." there is found no evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to complainant: filter stuck in the introducer and didn't advance normally.Doctor took it out and placed another celect through same introducer.Patient outcome: the patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key5902987
MDR Text Key52955111
Report Number3002808486-2016-00975
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529219
UDI-Public(01)10827002529219(17)150726(10)E2952990
Combination Product (y/n)N
Reporter Country CodeFI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-2-UNI-CELECT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/26/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight90
-
-