Merge eye station is intended to be used in conjunction with existing ophthalmic fundus cameras to take images of the eye, perform fluorescein angiography, red free, color and icg still-image photography as well as video imaging.On (b)(6) 2016, merge healthcare was notified a system recovery was occurring each time the eye station was turned on.On (b)(6) 2016, it was determined by merge healthcare support that the hard drives were corrupt.New hard drives were shipped to the customer.The hard drives were installed on (b)(6) 2016.On (b)(6) 2016, follow-up information was received from the customer.According to the customer, there were patients that were cancelled and rescheduled.On (b)(6) 2016, merge support configured the software and the issue was resolved.With merge eye station not working as expected there is a potential for a delay in diagnosis or treatment.However, there is no direct patient harm occurred as a result of the delay.Reference complaint number (b)(4).
|