MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K09-02334N
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/11/2016
Event Type  malfunction  
Manufacturer Narrative
One device was returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The user reported that the purchased drip chamber in their kit is running out of contrast before the bottle is empty.No air was injected into the patient.
 
Manufacturer Narrative
One device was returned for evaluation.The device was examined visually and simulated use testing was performed.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record and complaint database could not be reviewed as the lot of the returned unit was not confirmed.Corrective actions are in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer Contact
casey hughes, ms, cqe
1600 west merit parkway
south jordan, UT 84095
8013164932
MDR Report Key5913517
MDR Text Key53458623
Report Number1721504-2016-00169
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K913161
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberK09-02334N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-