• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE CEILING; SYSTEM, ANGIO, XRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY ARTIS ZEE CEILING; SYSTEM, ANGIO, XRAY Back to Search Results
Model Number 10094137
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/11/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that the artis zee ceiling system went down during a tavr procedure.All system movements and xray were halted.The patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The emergency stop circuit, when broken, will prohibit any movement of the stand and the table.It is activated by pressing one of the emergency buttons located on each control module and deactivated by releasing the button.Activation of the emergency stop can occur when a connection between the control modules is interrupted.The affected cable was exchanged and the problem did not reoccur.No further actions are to be taken as there is no negative awareness in regards to the quality and performance of the affected component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
SYSTEM, ANGIO, XRAY
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1
forchheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key5917502
MDR Text Key54346042
Report Number2240869-2016-51166
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K073290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 08/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10094137
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-