Currently, it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
The pump passed the displacement, self test, off no power, unexpected restart, rewind and basic occlusion tests.The pump was unable to prime during the prime test due to moisture damage on the force sensor resistor.Unable to perform the occlusion or excessive no delivery tests due to the prime/fill anomaly.Unable to confirm the no excessive no delivery alarms due to the prime/fill anomaly.All operating currents were within specification.The pump was received with a cracked reservoir tube lip, minor scratches on the display window, a cracked case at the display window corner, cracked battery tube threads and a cracked display window.
|