Catalog Number EMAX2PLUS
Device Problems Moisture Damage (1405); Overheating of Device (1437); Vibration (1674); Defective Device (2588); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Gtin is unavailable as the product made prior to compliance date; (b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device the reported condition was confirmed.It was determined that the issue was due to improper maintenance.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported by (b)(6) that the motor device had an undetermined malfunction.During the pre-repair diagnostics assessment, it was determined that the air pump was not functioning; the device was overheating, had excessive noise and vibration.It was further reported that the device was wet internally, which caused the motor to malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5933372
MDR Text Key54649827
Report Number1045834-2016-12655
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2016
Date Manufacturer Received08/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-