Reportable based on the analysis of the returned specimen.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested and was determined to be acceptable.Examination of the returned specimen revealed cuts by unknown means to the core wire 1.76cm from the proximal aspect of the proximal coil to core wire solder joint and 1.55cm from the proximal end of the distal coil segment; all material distal of the cuts is missing.The specimen also presents several bends of varying severity and areas of scraped ptfe coating scattered over the length of the device.Deposits of dried blood-like material are present on and between the coil wraps.No other damage or inconsistencies are noted to the specimen at this time.The proximal joint appears to be intact by visual examination and by non-destructive testing.Except where noted, the specimen device appears visually and dimensionally correct.Based on the evidence presented by the sample and the information provided by the supporting documentation, clinical and/or procedural factors appear to have impacted on the event as reported.
|