Model Number 9800
Device Problem Loss of Data (2903)
Patient Problem No Patient Involvement (2645)
Event Date 08/03/2016
Event Type  malfunction  
Manufacturer Narrative
A ge service representative performed an onsite investigation.The hard disk drive was evaluated and replaced.The system software and calibrations were also reloaded.The system was tested and found to be working as intended and returned to service.
 
Event Description
The customer reported that the system failed to properly save and recall patient image data.No patient serious injury or death was reported related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9800
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers drive
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers drive
salt lake city UT 84116
Manufacturer Contact
384 wright brothers drive
salt lake city, UT 84116
MDR Report Key5937911
MDR Text Key54267718
Report Number1720753-2016-02418
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9800
Device Lot Number89-2755
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse